×

E5070B-10B ENA Series Network

消耗积分:5 | 格式:rar | 大小:214 | 2010-08-21

旧念

分享资料个

Situation:
 When using the A1 source board (E5071-66571/69571) in repairs, Rev A.05.00 or later firmware is required. If
customer unit’s firmware version is less than Rev.A.05.00, it should be updated by changing the operating
system. With Rev A.03.60 or below version firmware, an error message “Power on test failed (A1 not found)” is
shown at lower left part on a display at power on. With Rev A.03.62 or A.04.00 version firmware, any of
messages  “Power on test failed (A1 DC-bus, 0x00000010)”, “Power on test failed (A1 DC-bus, 0x00000020)”,
or “Power on test failed (A1 DC-bus, 0x00000030)” is shown at power on.
 
To avoid troubling on customers, especially after a system recovery timing, it’s necessary to update the E5070B
operating system if customer unit’s firmware is Rev.A.4.00 or below.

声明:本文内容及配图由入驻作者撰写或者入驻合作网站授权转载。文章观点仅代表作者本人,不代表电子发烧友网立场。文章及其配图仅供工程师学习之用,如有内容侵权或者其他违规问题,请联系本站处理。 举报投诉

评论(0)
发评论

下载排行榜

全部0条评论

快来发表一下你的评论吧 !