×

关于ISO 80601-2-61 -2011之简介及性能分析

消耗积分:1 | 格式:pdf | 大小:1.29 MB | 2018-07-17

August辰

分享资料个

  Introduction

  The approximation of arterial haemoglobin saturation and pulse rate using pulse oximetry is common practice

  in many areas of medicine. This standard covers BASIC SAFETY and ESSENTIAL PERFORMANCE requirements

  achievable within the limits of existing technology.

  Annex AA contains a rationale for some of the requirements. It is included to provide additional insight into the

  reasoning of the committee that led to a requirement and identifying the HAZARDS that the requirement

  addresses.

  Annex BB is a literature survey relevant to the determination of the maximum safe temperature of the interface

  between a PULSE OXIMETER PROBE and a PATIENT‘S tissue.

  Annex CC discusses both the formulae used to evaluate the SpO2 ACCURACY of PULSE OXIMETER EQUIPMENT

  measurements, and the names that are assigned to those formulae.

  Annex DD presents guidance on when in vitro blood calibration of PULSE OXIMETER EQUIPMENT is needed.

  Annex EE presents a guideline for a CONTROLLED DESATURATION STUDY for the calibration of PULSE OXIMETER

  EQUIPMENT.

  Annex FF is a tutorial introduction to several kinds of testers used in pulse oximetry.

  Annex GG describes concepts of PULSE OXIMETER EQUIPMENT response time.

关于ISO 80601-2-61 -2011之简介及性能分析

声明:本文内容及配图由入驻作者撰写或者入驻合作网站授权转载。文章观点仅代表作者本人,不代表电子发烧友网立场。文章及其配图仅供工程师学习之用,如有内容侵权或者其他违规问题,请联系本站处理。 举报投诉

评论(0)
发评论

下载排行榜

全部0条评论

快来发表一下你的评论吧 !